G4 PHOENIX DH is mainly used for the determination of welding consumables and welding of diffusible hydrogen content, the instrument is equipped with infrared rapid heating furnace, and configure the samples got big foot diameter pipe for a sample of hydrogen. 10:00 inside the instrument integrates standard unit for correcting the air and using a high-sensitivity detection of long-term stable thermal conductivity detector for the diffusion of hydrogen, the instrument in full compliance with ISO3690 standards and the latest national standards.
Analysis speed, automation of welding samples in just 20 minutes
Can be rapidly heated to 900 ° C infrared furnace with temperature-programmed functions, from the trachea? 30mm
Optional maximum temperature of 1200 ° C tube furnace, from the trachea? 18mm
Can be adapted to analyze EN / ISO 3690 and AWS A4.3 standard large-size standard blocks
Long-term stability with high sensitivity thermal conductivity detector
Built-in 10-point standard calibration gas unit, no standard calibration
In line with ISO 3690 and AWS A4.3 data reporting format
Optional analysis result output format ppm or ml / 100g
By upgrading to achieve total diffusion of hydrogen and hydrogen detection in the same instrument
And mass spectrometry to realize ppb level content testing '
The results measured the diffusion of hydrogen and mercury can be a good match
Analysis speed, just 20 minutes, the mercury method required 72 hours
Automation, simple operation, the mercury method is complicated, the results need to be converted according to the outside temperature and pressure
Safe, no harmful gases or reagents, and mercury method because of the human body and the environment, has been banned by UNESCO
Measuring range:
0.05-1000ml / 100g
Analysis time:
Welding samples are generally 15-20 minutes
Depending on the weight of the sample material
Resolution:
0.001ml / 100g
Reproducibility:
± 0.01ml / 100g or ± 1% relative
Samples of different materials depending on weight
Carrier gas:
N2 minimum purity of 99.999%
Pressure 2bar
Cooling water (only infrared oven):
About 1l / min
Size and Weight: 460x750x500mm 50kg
Power supply:
When analyzing 230V, 1kVA
Standby 230V, 300VA
System Requirements:
Operating System: WindowsXP
CD-ROM
Two RS232 serial ports