JZB-1800 Syringe Pump

Loading Port:China Main Port

Payment Terms:TT OR LC

Min Order Qty:-

Supply Capability:-

Product Description:

Features


1. Compatible with syringes of most of the standard brands.
2. Automatic Calibration for syringes.
3. Three working modes: Rate control mode, Time control mode and Dose weight control mode

(ug/kg/min or mg/kg/h).
4. Built in rechargeable battery (Lithium-ion polymer battery).
5. Alarm functions: occlusion, empty, near empty, low battery, infusion over, syringe loose, etc
6. Three levels of the occlusion limited pressure: high, middle and low.
7. Three levels of alarm sound: high, middle and low.
8. The Syringe pump goes into KVO mode automatically as infusion is completed.
9. Purge / Bolus function.


Technical Specification


1. Flow range (Rate mode):
 50 ml syringe : 0.1 – 1800.0 ml / hr (0.1 ml / hr step)
 20 ml syringe : 0.1 – 800.0 ml / hr (0.1 ml / hr step)
 10 ml syringe : 0.1 – 400.0 ml / hr (0.1 ml / hr step)
2. Accuracy: +/- 2% (After correct calibration)
3. Time setting mode range:
 Volume: 0.1 – 199.9ml, 0.1ml step
 Time: 1 – 1999min, 1min step
4. Weight setting mode range:
 Sol. volume: 0.1 – 199.9ml, 0.1ml step
 Drug mass: 0.1 – 999.9ml, 0.1 ml step
 Dose: 0.01 – 99.99ug/kg/min, 0.01ug/kg/min step
0.01 – 99.99mg/kg/h, 0.01mg/kg/h step
Weight: 0.1 – 99.9kg, 0.1kg step
100 – 300kg, 1kg step
5. Purge rate:
 1800ml/h (50ml), 800ml/h (20ml), 400ml/h (10ml)
6. Volume limits range: 0 – 1999.9ml, 0.1ml step
7. Bolus volume: 1.0 – 100ml, 0.1ml step
8. Bolus Rate:
 10ml syringes: 1.0 – 400.0ml/hr, 0.1ml/hr step
 20ml syringes: 1.0 – 800.0ml/hr, 0.1ml/hr step
 50ml syringes: 1.0 – 1800.0 ml/hr, 0.1ml/hr step
9. Spec of the fuse: F1AL /250V, two (inside)
10. Power supply: AC100~240V, 50/60Hz, Internal Lithium-ion polymer battery: DC 11.1V, ≥1600mAh. The battery backup is approx. 8 hours (when using a fully charged battery and running at 5 ml/h).
11. Power consumption: ≤25VA
12. Class: I
13. Type: CF
14. Size: 280mm x 160mm x 130mm
15. Net Weight: 3.5 Kg. (Approx)
16. Operating condition:
 Ambient temperature: 10-40degree C
 Relative humidity: 20-90%
17. Storage condition:
 Ambient temperature: -10~55degree C
 Relative humidity: 10~90%.
18. Possibility of creation of a database about used solutions (drugs)